18
SWISS LITHOCLAST BLAST - FB-610/EN 23-JUNE-2017 / EMS CONFIDENTIAL
5. TREATMENT
To avoid the risk of electric shock, this product must
only be connected to a mains power supply with
protective earth. No modication shall be made on
this product. The mains power switch of the product
must be accessible at any time.
Do not use the product in surgery after any product
update without rst performing functional tests.
Fragments blocked in the lumen of the probe and
the handpiece may lead to loss of suction and
heating of the probe. If blockage occurs, stop litho-
tripsy. Use the unclogging rod to remove fragments
from the probe and from the handpiece lumen
before continuing.
Figure 49
Do not let the handpiece remain in contact with the
patient during treatment.
During treatment, an auditory information pulse
will be emitted.
This section provides guidance for using the product.
It does not provide detailed instructions for performing
lithotripsy procedures.
5.1. FUNCTIONAL TESTS
If a function or component is not working as
explained below, refer to the Troubleshooting
section.
1. From the STAND BY screen, press the START
button to access the READY screen.
2. Insert the probe into a sterile receptacle of physio-
logical uid.
3. Use the 2-mode foot pedal.
4. Press the pedal halfway (STEP 1) to activate suction
and make sure that suction is working properly (uid
moving through the suction tube).
1
Figure 50
5. Press the pedal completely (STEP 2) to activate both
suction and energies and make sure that the quality
meter is in the green zone and the uid is moving
through the suction tube.
1
2
Figure 51
6. Remove foot from the pedal to stop the functional
test.
Figure 52