14
SWISS LITHOCLAST BLAST - FB-610/EN 23-JUNE-2017 / EMS CONFIDENTIAL
2. Remove the protective cap from the console.
Figure 33
3. Connect the handpiece to the console.
Figure 34
4. Pay attention to the orientation of the handpiece
connector.
The red dot must be on top for proper alignment.
5. Make sure that the handpiece cord does not touch
the oor and is not compressed or squeezed in any
way that might impede circulation of the cooling liquid.
6. The handpiece connection to the console is maintained
by a mechanical lock. During use, the lock icon (orange
handpiece activation icon) remains illuminated.
Do not exceed the maximum number of usage cy-
cles for the handpiece as specied in the Techni-
cal Data section.
3.9. INSTALLING A PROBE ON THE HANDPIECE
1. Select the appropriate probe.
Risk of contamination: do not use after the expiration
date on the package label.
Refer to the Probe Compatibility Table section.
2. Use the wrench to rmly tighten the appropriate probe
on the handpiece.
Case 1: Standard wrench
Figure 35
Case 2: Multiuse Torque wrench
Figure 36
3.10. CONNECTING THE POWER CORD
Connect only to a FI protected mains power supply
(FI = Residual current protection).
To prevent damage to the console, make sure that
its rated voltage meets the local line voltage.
Connect the power cord to the power socket at the rear
of the console.
Figure 37