EasyManua.ls Logo

EMS Swiss LithoCast Trilogy FT-231 - 14 New Electromagnetic Compatibility

EMS Swiss LithoCast Trilogy FT-231
48 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
40
SWISS LITHOCLAST BLAST - FB-610/EN 23-JUNE-2017 / EMS CONFIDENTIAL
14. NEW ELECTROMAGNETIC COMPATIBILITY
Electromagnetic compatibility according to IEC 60601-1-2:2014
Guidance and manufacturers declaration – electromagnetic emissions
The Swiss LithoClast® Trilogy is intended for use in the electromagnetic environment specied below. The
customer or the user of the Swiss LithoClast® Trilogy should assure that it is used in such an environment.
EMISSIONS TEST COMPLIANCE ELECTROMAGNETIC ENVIRONMENT - GUIDANCE
RF emissions
CISPR 11
Group 1
The Swiss LithoClast® Trilogy uses RF energy only for its internal
function. Therefore, its RF emissions are very low and are not
likely to cause any interference in nearby electronic equipment.
RF emissions
CISPR 11
Class A
The emissions characteristics of the Swiss LithoClast® Trilogy
make it suitable for use in hospitals only.
Harmonic emissions
IEC 61000-3-2
Not applicable
Voltage uctuation / icker
emissions
IEC 61000-3-3
Not applicable
Table 13
Guidance and manufacturers declaration – electromagnetic immunity
The Swiss LithoClast® Trilogy is intended for use in the electromagnetic environment specied below. The
customer or the user of the Swiss LithoClast® Trilogy should assure that it is used in such an environment.
IMMUNITY TEST
IEC 60601
TEST LEVEL
COMPLIANCE LEVEL
ELECTROMAGNETIC
ENVIRONMENT - GUIDANCE
Electrostatic discharge
(ESD)
IEC 61000-4-2
± 8 kV contact
± 15 kV air
± 8 kV contact
± 15 kV air
Floors should be wood, concrete
or ceramic tile. If oors are
covered with synthetic material,
the relative humidity should be at
least 30%.
Radiated RF
IEC 61000-4-3
3 V/m
80 MHz to 2.7 GHz
10 V/m
80 MHz to 2.7 GHz
Portable and mobile RF communi-
cations equipment should be used
no closer than 30 cm to any part
of the Swiss LithoClast® Trilogy,
including cables.
Proximity elds from RF
wireless communica-
tions equipment
IEC 61000-4-3
See next table See next table
Conducted RF
IEC 61000-4-6
3 V rms
150 kHz to 80 MHz
6 V rms
in ISM and amateur radio
bands
3 V rms
150 kHz to 80 MHz
6 V rms
in ISM and amateur radio
bands
Power frequency
(50/60 Hz) magnetic
eld
IEC 61000-4-8
30 A/m 30 A/m
Power frequency magnetic elds
should be at levels character-
istic of a typical location in a
typical commercial or hospital
environment.
Table 14
UT is the A/C mains voltage prior to application of the test level.