2. If required, adjust any settings manually as described in the following table:
PICTOGRAMS MEANING ACTION
ON/OFF button
Use the ON/OFF button to activate or deactivate the functionality in
question.
Impact
Impact power
Use the
and buttons to adjust the impact power in percent
from 10% to 100% (in 10% increments).
Impact frequency
Use the
and buttons to adjust the frequency of impact
pulses from 1 Hz to 12 Hz (in 1 Hz increments).
Ultrasound
Ultrasound power
Use the and buttons to adjust the ultrasound power from
10% to 100% (in 10% increments).
Suction
Suction ow rate
Use the and buttons to adjust the suction ow rate from
10% to 100% (in 10% increments).
This control is only active for consoles with an integrated
peristaltic pump device.
Treatment Efciency
Indicator
To provide instant visual feedback about the efciency of the
treatment.
• Green: the treatment works properly.
• Orange: the treatment is not efcient.
Menu To return to the STAND BY screen from the READY screen.
Table 3
5.3.2. Hard Stones Treatment Settings
1. To use the hard stones pre-settings, press the HARD
STONES TREATMENT button from the STAND BY
screen.
Figure 56
2. The READY screen will appear and display the hard
stone treatment pre-settings.
3. If required, adjust any settings manually as described
in the table above.
5.3.3. Soft Stones Treatment Settings
1. To use the soft stones pre-settings, press the SOFT
STONES TREATMENT button from the STAND BY
screen.
2. The READY screen will appear and display the soft
stones treatment pre-settings.
3. If required, adjust any settings manually as described
in the table above.