XA02712P-A Cerabar M PMC51, PMP51, PMP55
10 Endress+Hauser
Device type PMP51, PMP55
Maximum surface
temperature
Process temperature range Ambient temperature range
T125 °C –40 °C ≤ T
p
≤ +80 °C –40 °C ≤ T
a
≤ +60 °C
–40 °C ≤ T
p
≤ +100 °C –40 °C ≤ T
a
≤ +55 °C
–40 °C ≤ T
p
≤ +125 °C –40 °C ≤ T
a
≤ +45 °C
Basic specification, Position 16 = B, D, E
Maximum surface
temperature
Process temperature range Ambient temperature range
T125 °C –40 °C ≤ T
p
≤ +200 °C –40 °C ≤ T
a
≤ +65 °C
–40 °C ≤ T
p
≤ +300 °C –40 °C ≤ T
a
≤ +60 °C
–40 °C ≤ T
p
≤ +400 °C –40 °C ≤ T
a
≤ +55 °C
Specific conditions of use:
The surface temperature is
for equipment protection level (EPL) Da: T
200
125 °C (with 200 mm
dust deposit)
and equipment protection level (EPL) Db: T
L
125 °C (with dust
accumulation T
L
)
T
L
marking:
The assigned surface temperature without dust layer is the same.
Connection data
Basic specification, Position 3 = 2
Power supply
U ≤ 45 V
DC
Basic specification, Position 3 = 3, 4
Power supply
U ≤ 32 V
DC