EasyManua.ls Logo

Endress+Hauser Cerabar M PMC51 - Page 11

Endress+Hauser Cerabar M PMC51
12 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Cerabar M PMC51, PMP51, PMP55 XA02712P-A
Endress+Hauser 11
Cable entry: Connection compartment
Ex tb
Thread Clamping range Material Sealing insert O-ring
M20x1,5 ø 8 to 10.5 mm
1)
ø 6.5 to 13 mm
2)
Ms, nickel-plated Silicone EPDM (ø 17x2)
1) Standard
2) Separate clamping inserts available
The tightening torque refers to cable glands installed by the
manufacturer:
Recommended: 3.5 Nm
Maximum: 10 Nm
This value may be different depending on the type of cable.
However, the maximum value must not be exceeded.
Only suitable for fixed installation. The operator must pay attention
to a suitable strain relief of the cable.
The cable glands are suitable for a low risk of mechanical danger
(4 Joule) and must be mounted in a protected position if larger
impact energy levels are expected.
To maintain the ingress protection of the enclosure: Install the
enclosure cover, cable glands and blind plugs correctly.

Other manuals for Endress+Hauser Cerabar M PMC51

Related product manuals