EasyManua.ls Logo

Endress+Hauser cerabar S PMC Series - Page 37

Endress+Hauser cerabar S PMC Series
68 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Manual entry
Requirements for manual linearisation are
as follows:
The max. 21 pairs of values for the
points on the linearisation curve are
known.
The curve is given as a % level
(% pressure span) against % volume.
The linearisation curve must rise or fall
continuously.
The measured value is supplied as a
volume.
Note!
For step 1, you can also carry out a
zero correction according to the
procedure described in Chapter 5.1,
page 27.
A calibration can also be made with a
reference pressure using steps 1-3,
see page 34.
In edit mode V3H6 = Manual, you can
delete individual points in a linearisation
table entering "9999" for level or
volume. But first activate the
linearisation table.
Entering the pairs of values for the
linearisation curve is done after a
calibration with reference pressure or a dry
calibration in %. The procedure for dry
calibration is described below (see also
page 35).
# Matrix Path through the
menus
Entry
1 Set the display to "0"
V0H5 Sets
bias pressure
e.g. 0.1 mbar
Confirm
2 Enter the theoretical pressure for the lower
range-value (empty)
V0H1 Sets 4 mA 0 mbar
Confirm
3 Enter the theoretical pressure for the upper
range-value (full)
V0H2 Sets 20 mA 500 mbar
Confirm
# Changing product?
see "Density Correction" page 33
Main group: Linearisation
4 Select linearisation mode "Manual entry"
V3H6 Operation mode
manual Confirm
5 Enter the table
V3H7 Line No. 1
Confirm
V3H8 Input level e.g. 0 %
Confirm
V3H9 Set volume e.g. 0 %
Confirm
Repeat step 5 until all points have been
entered.
6 Activate the table
V3H6 Activate
table Confirm
7 Select operation mode "Level manual"
V3H0 Level
manual Confirm
8 Enter the level or volume at the minimum level
V3H1 Display at
4 mA
e.g. 0 (hl)
Confirm
9 Enter the level or volume at the maximum level
V3H2 Display at
20 mA
e.g. 10 (hl)
Confirm
10 Select the level or volume units
(Select unit from the tables on 32)
V3H3 Unit after
linearisation
e.g. hl
Confirm
Note!
1
4
6
5
3
2
7
BA187Y56
100 %
0 %
Level
V
Empty
V
Full
Volume
Table giving example
Point Measured
value
(mbar)
Level
(%)
Volume
(%)
1000
2 100 20 8
3 200 40 20
7 500 100 100
volume at x % level =
total volume volume (%)
10 0
Cerabar S 6 Level Measurement
Endress+Hauser 37

Related product manuals