Operation Ceraphant T PTC31, PTP31, PTP35
12 Endress+Hauser
5.1.1 Navigating in the operating menu
P01-PTx3xxxx-19-xx-xx-xx-005
Fig. 7: Navigating in the operating menu
A Function group selection
B Function selection
Enter the operating menu
– press the E key for longer than 3 s
Select the "Function group" with the + or - key
Select the "Function" with the E key
Enter or change parameters with the + or - key
then return to "Function" with the E key Note: If software locking is enabled, it must be disabled before making
entries or changes by entering the code selected
Press the E key several times to return to the "Function group"
Jump back to the measuring position (Home position)
– press the E key for longer than 3 s
Query to save data (select "YES" or "NO" with the + or - key)
– confirm with the E key
!
Note!
Changes to the parameter settings only become effective if you choose s "YES" when asked to
save data.