Deltabar FMD72 Commissioning
Endress+Hauser 31
10.5.2 Configuring language via operating tool (FieldCare)
See the Operating Instructions BA01044P on the CD-ROM provided.
10.5.3 Measuring mode selection
Measuring mode (005)
Navigation
Setup ® Measuring mode
Write permission
Operators/Service engineers/Expert
Description
Select the measuring mode.
The operating menu is structured differently depending on
the measuring mode selected.
Options
Pressure
Level
Factory setting
Level
Note
If the measuring mode is changed, no conversion takes
place. If necessary, the device has to be recalibrated after
the measuring mode has been changed.
10.5.4 Defining the high-pressure side
High Press. Side (183)
Navigation
Setup ® High Press. Side
Write permission
Operators/Service engineers/Expert
Description
Define which sensor module corresponds to the high-
pressure side.
Options
Sensor module HP
Sensor module LP
Factory setting
Sensor module HP