Commissioning Deltabar FMD72
32 Endress+Hauser
10.5.5 Press. Eng. Unit selection
Press. Eng. Unit (125)
Navigation
Setup ® Press. Eng. Unit
Write permission
Operators/Service engineers/Expert
Description
Select the pressure engineering unit. If a new pressure
engineering unit is selected, all pressure-specific
parameters are converted and displayed with the new unit.
Options
mbar, bar
mmH2O, mH2O,
in H2O, ftH2O
Pa, kPa, MPa
psi
mmHg, inHg
kgf/cm
2
Factory setting
mbar, bar or psi depending on the sensor module nominal
measuring range, or as per order specifications
10.6 Pos. Zero Adjust
The pressure resulting from the orientation of the device can be corrected here.
Corrected press. (172)
Navigation
Setup ® Corrected Press.
Write permission
Operators/Service engineers/Expert
Description
Displays the measured pressure after the differential
pressure buildup and position adjustment.
Note
If this value is not equal to "0", it can be corrected to "0"
by the position adjustment.
Pos. Zero Adjust (007)