Deltabar FMD72 Commissioning
Endress+Hauser 33
Navigation
Setup ® Pos. Zero Adjust
Write permission
Operators/Service engineers/Expert
Description
Position adjustment the pressure difference between zero
(set point) and the measured differential pressure need not
be known.
Options
Confirm
Cancel
Example
Measured value = 2.2 mbar (0.033 psi)
You correct the measured value via the "Pos. Zero
Adjust" parameter with the "Confirm" option. This
means that you are assigning the value 0.0 to the
pressure present.
Measured value (after pos. zero adjust) = 0.0 mbar
The current value is also corrected.
Factory setting
Cancel
10.7 Configuring level measurement
10.7.1 Information on level measurement
You have a choice of two methods for calculating the level: "In pressure" and "In
height". The table in the "Overview of level measurement" section that follows
provides you with an overview of these two measuring tasks.
The limit values are not checked, i.e. the values entered must be appropriate for the
sensor module and the measuring task for the device to be able to measure correctly.
Customer-specific units are not possible.
There is no unit conversion.
The values entered for "Empty Calib./Full Calib.", "Empty Pressure/Full Pressure",
"Empty Height/Full Height", and "Set LRV/Set URV" must be at least 1% apart. The
value will be rejected, and a warning message displayed, if the values are too close
together.