Electrical connection Deltabar FMD72
16 Endress+Hauser
6
7
5
5
6
2 3 4
3
4
1
21
6
NPT ½"
M20
8
A0017528
1 BK (black)
2 BU (blue)
3 WH (white)
4 BN (brown)
5 Sensor module LP
6 Sensor module HP
7 Ground terminal
8 Torque 0.4 Nm
7.1.1 Screening with cable shield
Screening with cable shield is described in the associated documentation SD00354P. The
documentation is provided with the connecting cables.
7.2 Connecting the sensor module HP to the transmitter
!
WARNING
Is the supply voltage switched off?
Risk of electric shock and/or explosion!
►
Switch off the supply voltage before connecting the device.