Deltapilot M 4...20 mA HART Commissioning
Endress+Hauser 41
The measured variables %, level, volume and mass are available for this level mode. See
operating instructions "Unit before lin (025)".
8.5.4 "In pressure" level selection
Calibration without reference pressure (dry calibration)
Example:
In this example, the volume in a tank should be measured in liters. The maximum volume of
1000 liters (264 US gal) corresponds to a pressure of 450 mbar (6.75 psi). The minimum
volume of 0 liters corresponds to a pressure of 50 mbar (0.75 psi) since the device is mounted
below the start of the level measuring range.
Prerequisite:
The measured variable is in direct proportion to the pressure.
9 The pressure for the upper calibration point is
present at the device, here 300 mbar (4.5 psi) for
example.
A0031063
Fig. 11: Calibration with reference pressure –
wet calibration
A See Table, Step 8.
B See Table, Step 9.
C See Table, Step 10.
D See Table, Step 11.
Select the "Full calib." parameter.
Menu path: Setup Extended setup Level
Full calib.
Enter the level value, here 3 m (9.8 ft) for example.
The pressure value present is assigned to the upper
level value by confirming the value.
10 Set the level value for the lower current value (4
mA) by means of "Set LRV".
Menu path: Setup Extended setup Current
output Set LRV
11 Set the level value for the upper current value (20
mA) by means of "Set URV".
Menu path: Setup Extended setup Current
output Set URV
12 If calibration was performed with a medium other
than the process medium, specify the density of the
process medium in the "Process density" parameter.
Menu path: Setup Extended setup Level
Process density
13 Result:
The measuring range is set for 0 to 3 m (9.8 ft).
Description
C
D
I
[mA]
h
[m]
20
4
0 3