Commissioning Deltapilot M 4...20 mA HART
48 Endress+Hauser
• The values for "Empty calib./Full calib.", "Empty height/Full height" and" Set LRV/Set URV"
must be at least 1% apart. The value will be rejected, and a warning message displayed, if the
values are too close together. Further limit values are not checked, i.e. the values entered must
be appropriate for the sensor and the measuring task for the measuring device to be able to
measure correctly.
• Due to the orientation of the device, there may be pressure shifts in the measured value, i.e.
when the container is empty or partly filled, the measured value is not zero. For information
on how to perform position adjustment, see ä 37, "Position zero adjustment".
Description
1 Select the "Level" measuring mode via the
"Measuring mode" parameter.
Menu path: Setup Measuring mode
A0030030
Fig. 16: Calibration without reference pressure –
dry calibration
A See Table, Step 11.
B See Table, Steps 7 and 8.
D See Table, Steps 9 and 10.
2 Select a pressure unit via the "Press eng. unit"
parameter, here "mbar" for example.
Menu path: Setup Press. eng. unit
3 Select the "In height" level mode via the "Level
selection" parameter.
Menu path: Setup Extended setup Level
Level selection
4 Select a level unit via the "Unit before lin" parameter,
here "l" (liters) for example.
Menu path: Setup Extended setup Level
Unit before lin
5 Select a level unit via the "Height unit" parameter,
here "m" for example.
Menu path: Setup Extended setup Level
Height unit
6 Select the "Dry" option via the "Calibration mode"
parameter.
Menu path: Setup Extended setup Level
Calibration mode
7 Enter the volume value for the lower calibration
point via the "Empty calib." parameter, here 0 liters
for example.
Menu path: Setup Extended setup Level
Empty calib.
B
1000 l
450 mbar
A
0 l
50 mbar