Gammapilot M FMG60 XA00303F-G
Endress+Hauser 17
of a number or a letter (e.g. J = Test, Certificate). The second digit constitutes the value that stands
for the feature within the group (e.g. A = 3.1 material (wetted parts), inspection certificate).
More detailed information about the device is provided in the following tables. These tables describe
the individual positions and IDs in the extended order code which are relevant to hazardous
locations.
Extended order code: Gammapilot M
Device type
FMG60
Basic specifications
Position 1 (Approval)
Selected option Description
FMG60 1 ATEX II 2(1) G Ex db eb [ia Ga] IIC T6 Gb
2 ATEX II 2(1) G Ex db eb [ia Ga] IIC T6 Gb, WHG
3 ATEX II 2(1) G Ex db [ia Ga] IIC T6 Gb
4 ATEX II 2(1) G Ex db [ia Ga] IIC T6 Gb, WHG
6 ATEX II 2(1) G Ex db eb [ia Ga] IIC T6 Gb
ATEX II 2(1) D Ex tb [ia Da] IIIC T80°C/T85°C Db
7 ATEX II 2(1) G Ex db eb [ia Ga] IIC T6 Gb, WHG
ATEX II 2(1) D Ex tb [ia Da] IIIC T80°C/T85°C Db
8 ATEX II 2(1) G Ex db [ia Ga] IIC T6 Gb
ATEX II 2(1) D Ex tb [ia Da] IIIC T80°C/T85°C Db
M ATEX II 2(1) G Ex db [ia Ga] IIC T6 Gb, WHG
ATEX II 2(1) D Ex tb [ia Da] IIIC T80°C/T85°C Db
Position 2 (Power Supply)
Selected option Description
FMG60 1 90-253 VAC
2 18-35 VDC
Position 3 (Connect. Power Supply; Connect. Output)
Selected option Description
FMG60 B Ex eb; Ex ia
C Ex eb; Ex eb
D Ex db (XP); Ex db (XP)
E Ex db (XP); Ex ia (IS)
G Ex eb, dust-Ex; Ex eb, dust-Ex
H Ex db, dust-Ex; Ex db, dust-Ex
J Ex eb, dust-Ex; Ex ia, dust-Ex
K Ex db, dust-Ex; Ex ia, dust-Ex
Position 4 (Output)
Selected option Description
FMG60 1 4-20 mA HART