XA00303F-G Gammapilot M FMG60
22 Endress+Hauser
Basic specification, Position 5 (Scintillator; Measuring Range) = A-D, G-T
Ambient temperature
Detector without water cooling or detector with water cooling
out of operation:
Temperature class T6
• Devices with NaJ crystal scintillator:
• Devices with PVT plastic scintillator:
–40 °C ≤ T
a
≤ +60 °C
–40 °C ≤ T
a
≤ +60 °C
Detector with water cooling in operation: Temperature class T6 or T5
At the pipe housing (within the water cooling):
• Devices with NaJ crystal scintillator:
• Devices with PVT plastic scintillator:
–40 °C ≤ T
a
≤ +60 °C
–40 °C ≤ T
a
≤ +60 °C
At the compartment housing: –40 °C ≤ T
a
≤ +75 °C (T6)
–40 °C ≤ T
a
≤ +80 °C (T5)
Type of protection
Power supply circuit (Terminal compartment A) Ex eb or Ex db
Signal circuits (Terminal compartment B) Ex ia
Connection data
• Tightening torque of the terminal screws: max. 0.4 Nm.
• Strip the insulation of the connection wires with suitable length. Bare parts of the wires must not
emerge from the terminal.
• Ensure that the wires are securely clamped.
Terminal compartment A
Increased safety (Ex eb) or Flameproof enclosure (Ex db)
A0031763
9
Power supply circuit
Basic specification, Position 2
(Power Supply) = 1
L1
N
U
e
= 90 to 253 V
AC
, 50/60 Hz, 8.5 VA
Basic specification, Position 2
(Power Supply) = 2
L+
L–
U
e
= 18 to 35 V
DC
, 3.5 W
U
m
= 253 V
AC