Gammapilot M FMG60 XA00303F-G
Endress+Hauser 23
Signal circuit
Basic specification, Position 3
(Connect. Power Supply;
Connect. Output) = B, E, J, K
+ -
not connected
Basic specification, Position 3
(Connect. Power Supply;
Connect. Output) = C, D, G, H
+ -
4...20 mA/HART (active)
U
m
= 253 V
AC
The detector ensures galvanic isolation up to a maximum of
253 V
AC
between the signal circuit and any other circuit.
Terminal compartment B
Intrinsic safety (Ex ia)
+ –
IN CASCADE OUT
SIM WP
OFF
ON
3 4
5 6
7
8
9 10
1112
PT100
+ – – +
A0031764
10
Signal circuit Ex ia IIC Ex ia IIB Ex ib IIC Ex ib IIB
4...20 mA/ HART
(active)
1)
+ -
U
o
= 21.2 V
I
o
= 92 mA
P
o
= 479 mW
R
i
= 235 Ω
C
i
= 13.4 nF
L
i
= 0
Characteristic
curve: linear
U
i
= 30 V
I
i
= 13 mA
P
i
= 390 mW
C
o
= 150 nF
at
L
o
= 0.15 mH
C
o
= 115 nF
at
L
o
= 1 mH
C
o
= 886 nF
at
L
o
= 0.15 mH
C
o
= 686 nF
at
L
o
= 1 mH
C
o
= 169 nF
L
o
= 4 mH
C
o
= 1.2 µF
L
o
= 15 mH
PT100
U
o
= 8.4 V
I
o
= 8.3 mA
P
o
= 17.5 mW
R
i
= 1 012 Ω
C
i
= 0
L
i
= 0
Characteristic
curve: linear
C
o
= 1 200 nF
at
L
o
= 1 mH
C
o
= 1 800 nF
at
L
o
= 0.15 mH
C
o
= 6 µF
at
L
o
= 1 mH
C
o
= 5.2 µF
at
L
o
= 2 mH
C
o
= 5.2 µF
L
o
= 400 mH
C
o
= 43 µF
L
o
= 400 mH
Cascade out
2)
- +
U
o
= 8.4 V
I
o
= 19.2 mA
P
o
= 40.3 mW
R
i
= 439 Ω
C
i
= 5.3 nF
L
i
= 67 µH
Characteristic
curve: linear
C
o
= 5.1 µF
L
o
= 69 mH
C
o
= 42 µF
L
o
= 199 mH
C
o
= 5.1 µF
L
o
= 69 mH
C
o
= 42 µF
L
o
= 199 mH
Cascade in
3)
+ -
U
i
= 8.4 V
I
i
= 19.2 mA
P
i
= 40.3 mW
C
i
= 0
L
i
= 67 µH