XA00303F-G Gammapilot M FMG60
24 Endress+Hauser
Signal circuit Ex ia IIC Ex ia IIB Ex ib IIC Ex ib IIB
Connection for
FHX40
U
o
= 4.7 V
I
o
= 37.7 mA
P
o
= 44.3 mW
R
i
= 125 Ω
C
i
= 12.7 nF
L
i
= 0
Characteristic
curve: linear
For connection to the approved display FHX40 with
associated cable (KEMA 02 ATEX 1203) in type of protection
intrinsic safety Ex ia IIC or Ex ia IIIC.
Observe associated Safety Instructions XA00193F!
C
o
= 150 µF
L
o
= 25 mH
This circuit may also be connected to the approved Service Interface FXA193
with associated connection cable for ToF devices (BVS 03 ATEX E187).
Observe associated Safety Instructions XA00077D!
Output FXA193 with ToF-Cable:
U
o
= 3.74 V, I
o
= 9.9 mA, P
o
= 9.2 mW, C
i
= 0, L
i
= 0;
Characteristic curve: linear
For explosion group IIC: L
o
< 340 mH, C
o
< 100 µF
1) Only available at Basic specification, Position 3 (Connect. Power Supply; Connect. Output) = B, E, J, K
2) Only for connection to FMG60 signal circuit Cascade in
3) Only for connection to FMG60 signal circuit Cascade out