TMT162 Commissioning
Endress+Hauser 27
!
Note!
In order to activate the values to be shown in the device display using Commuwin II
and HART
®
handheld module DXR275/375: Add (DXR,CW=x) of the values to be
displayed and enter the sum.
• Display: time (2s, 4s, 6s, 8s)
• Display: figures after decimal point (0,1,2)
• Display PV text (customer specific text, 8 characters)
≥ SW 01.03.00
≥ SW 01.03.00
+
≥ SW 01.03.00
≥ SW 01.03.00
+
-
-
V6H1
Function group DISPLAY
Available in ReadWin
®
2000, HART
®
handheld module DXR275/375 (Symbol
7
),
Commuwin II with Matrix address
ReadWin
®
2000
7
/FieldCare CW II-
Matrix
Function group DIAGNOSTICS
Available in ReadWin
®
2000, HART
®
handheld module DXR275/375 (Symbol
7
),
Commuwin II with Matrix address
ReadWin
®
2000
7
/FieldCare CW II-
Matrix
Diagnostics Display of information required for device diagnostics.
• Device status or error code
(See chapter 9.2 "Error messages")
• Last error code (status) or previous error code
(See chapter 9.2 "Error messages")
• Status sensor 1 (0 = no error; 0 ≠ error)
• Status sensor 2 (0 = no error; 0 ≠ error)
• Configuration changed
+
+
-
-
+
+
+
+
+
+
•V9H0
•V9H1
•V0H4
•V0H6
•V9H2
Diagnostics •Static revision
The "Static revision" is increased on every parameter change. This is for
compliance to 21 CFR Part 11, showing that no further parameter changes
have been made.
•Sensor 1 max. value
• Sensor 1 min. value
•Sensor 2 max. value
• Sensor 2 min. value
•RJ max. value
•RJ min. value
Display of the maximum process value. The process value will be accepted after
starting the measurement.
Display of the minimum process value. The process value will be accepted after
starting the measurement.
Display of the maximum and minimum measured temperatures of the internal
Pt100 DIN B comparison measurement point.
!
Note!
Maximum process value is changed to the actual process value on write access.
On reset to factory default value the default value is entered -10000.
Minimum process value is changed to the actual process value on write access.
On reset to factory default value the default value is entered +10000.
-
+
+
+
+
+
+
-
+
+
+
+
+
+
•V9H3
•V8H0
•V8H1
•V8H2
•V8H3
•V8H4
•V8H5
Function group IDENTIFICATION
Available in ReadWin
®
2000, HART
®
handheld module DXR275/375 (Symbol
7
),
Commuwin II with Matrix address
ReadWin
®
2000
7
/FieldCare CW II-
Matrix
Measuring point
Input and display of the information relating to the measuring point identification