Commissioning TMT162
28 Endress+Hauser
Measuring
point TAG
Input: 8 characters + + VAH0
Description Input: 16 characters + + VAH1
Message Input: 32 characters - +
Device information
Display of the information relating to the device identification
Commuwin
device version
Special Commuwin device version, e.g.: 8010 means Version 1.0 - - VAH3
Device release Display of device release - + VAH2
Serial number 11 digit display of the Endress+Hauser device serial number
(equal to that on the legend plate).
++VAH4
Software
version
Display of the software version + + VAH6
Hardware
version
Display of the hardware version + + VAH7
Certificates Display of device approvals - +
Device
Display of the information relating to the HART
®
device identification
Manufacturer Manufacturer’s identification character: Endress+Hauser (=17) - + -
Device type Device type identification: TMT162 - + -
Date Individual use of this parameter - + -
Hardware
revision
Revisions of the device’s electronic components - + -
Function group IDENTIFICATION
Available in ReadWin
®
2000, HART
®
handheld module DXR275/375 (Symbol
7
),
Commuwin II with Matrix address
ReadWin
®
2000
7
/FieldCare CW II-
Matrix
Function group SERVICE FUNCTIONS
Available in ReadWin
®
2000, HART
®
handheld module DXR 275/375 (Symbol
7
),
Commuwin II with Matrix address
ReadWin
®
2000
7
/FieldCare CW II-
Matrix
Security
locking
Set-up release code.
Input:
• Lock = 0
• Release (unlock) = 261
++V9H6
Reset to default Reset to factory default values.
Input: 162
0
++V9H5