Installation iTEMP TMT71, TMT72
12 Endress+Hauser
100 (3.94)
12.5 (0.49)
H
116 (4.57)
A0039296
H The height of housing H varies depending on the terminal version: screw terminals =
114 mm (4.49 in), push-in terminals = 111.5 mm (4.39 in)
4.1.2 Mounting location
Head transmitter:
In the terminal head, flat face, as per DIN EN 50446, direct mounting on insert with cable
entry (middle hole 7 mm)
In the field housing, separated from the process
With DIN rail clip on DIN rail as per IEC 60715, TH35
DIN rail transmitter:
In DIN rail housing on DIN rail as per IEC 60715, TH35
NOTICE
When using DIN rail transmitters with a thermocouple/mV measurement, increased
measurement deviations may occur depending on the installation situation and ambient
conditions.
If the DIN rail transmitter is mounted on the DIN rail without any adjacent devices, this
may result in deviations of up to ± 1.34 °C. If the DIN rail transmitter is mounted in series
between other DIN rail devices (reference operating conditions: 24 V, 12 mA), deviations
of up to + 2.94 °C may occur.