iTEMP TMT71, TMT72 Installation
Endress+Hauser 17
4.2.2 Mounting the DIN rail transmitter
NOTICE
Wrong orientation
Measurement deviates from the maximum accuracy rating when a thermocouple is connected
and the internal reference junction is used.
Mount the device vertically and ensure it is oriented correctly!
1.
2.
3.
A0039678
 7 Mounting the DIN rail transmitter
1. Position the top DIN rail groove at the top end of the DIN rail.
2. Slide the bottom of the device over the bottom end of the DIN rail until you can hear the
lower DIN rail clip click into place on the DIN rail.
3. Pull gently on the device to check if it is correctly mounted on the DIN rail.
If it doesn't move, the DIN rail transmitter is correctly mounted.
4.3 Post-installation check
After installing the device, always run the following final checks:
Device condition and specifications Notes
Is the device undamaged (visual inspection)? -
Do the ambient conditions match the device specification (e.g. ambient temperature,
measuring range, etc.)?
→  13