Levelflex FMP53 FOUNDATION Fieldbus Mounting
Endress+Hauser 13
6.2 Mounting the device
6.2.1 Required mounting tools
To shorten rod or coax probes: Saw
For flanges and other process connections: appropriate mounting tools
To turn the housing: Hexagonal wrench 8 mm
6.2.2 Mounting the "Sensor remote" version
This section is only valid for devices of the version "Probe Design" = "Sensor remote"
(feature 600, option MB or MC).
For the version "Probe design" = "Sensor remote" the following is supplied:
The probe with the process connection and the connection cable (3m/9ft or 6m/18ft)
The electronics housing
The mounting bracket for wall or pipe mounting of the electronics housing
The connection cable is fixed to the probe on delivery.
The probe with connection cable and the electronics are adjusted to match each other.
They are marked by a common serial number. Only components with the same serial
number shall be connected to each other.
L
CAUTION
Mechanical stress may damage the plugs of the connection cable or cause accidental
loosening of the plugs.
Mount the probe and the electronics housing tightly before connecting the cable.
Lay the cable such that it is not exposed to mechanical stress. Minimum bending radius: 50
mm (2").
Torque for the coupling nut at the elctronics housing: 6 Nm
Torque for the coupling nut at the probe: 20 Nm
If the meeasuring point is exposed to strong vibrations, an additional locking compound
(e.g. Loctite 243) can be applied at the plug connector of the electronics housing.