Levelflex M FMP40
Endress+Hauser 25
Installation in concrete silos
Installation, for example, into a thick concrete ceiling should be made flush with the lower edge. Alternatively,
the probe can also be installed into a pipe that must not protrude over the lower edge of the silo ceiling.
The pipe should kept at a minimum length. Installation suggestions see diagram.
L00-FMP4xxxx-17-00-00-en-008
Note for installations with rod extension/center washer (accesories):
Strong dust generation can lead to build-up behind the center washer.
This can cause an interference signal. For other installation possibilities please contact Endress+Hauser.
Installation from the side
• If installation from above is not possible, the
Levelflex can also be mounted from the side.
• In this case, always fix the rope probe ("Fixing rope
probe", → ä 26).
• Support rod and coax probe if the lateral load-
bearing capacity is exceeded (→ ä 6).
Only fix rod probes at the probe end.
Caution!
Remove or ground the electronics when welding the
sleeve as the device will otherwise be destroyed!
L00-FMP4xxxx-17-00-00-en-035