Levelflex M FMP40
Endress+Hauser 29
Measurement in corrosive liquids
For measurement in corrosive liquids use Levelflex M FMP41C. When using plastic tanks it is also possible to
mount the probe on the outside of the tank (see installation instructions, ä 30).
Levelflex measures the level of aqueous media through the plastic.
Installation in underground tanks
Use coax probe for nozzles with large diameters in
order to avoid reflections at the nozzle wall.
L00-FMP4xxxx-17-00-00-yy-022
Installation in stilling well or bypass
A rod probe can be used for pipe diameters bigger
than 40 mm.
When installing a rod probe into a metallic pipe
with internal diameter of up to 150 mm, you have
all the advantages of a coax probe.
Welded joints that protrude up to approx. 5 mm
inwards do not influence measurement.
If a rod probe is used, the probe length must be
100 mm longer than the lower disposal.
It must be ensured that the probe does not come
into contact with the side wall. Where necessary,
use a centering disk at the lower end of the probe
( ä 61, "Type of Probe:")
L00-FMP4xxxx-17-00-00-yy-023
min. 100