Liquicap T FMI21
10 Endress+Hauser
Mechanical construction
!
Note!
All dimensions in mm (100 mm = 3.94 in)
Design, dimensions Rod probe
Process connection with parallel thread G 1½ or with tapered thread NPT 1½.
L00-FMI21xxx-06-05-xx-en-000
* Ground rod (uninsulated)
** Probe rod insulated
Probe shortening set
L00-FMI21xxxx-17-06-xx-en-000
!
Note!
If shortening the probe rods, the insulation above the cut must not be damaged!
Recalibration is always necessary after shortening the probe. The active rod probe begins above the probe
shortening set. Therefore the zero balance should be carried out at this place. Only the insulated rod must be
insulated with the shortening set. The ground rod remains uninsulated.