EasyManua.ls Logo

Endress+Hauser Liquicap T FMI21 - Operating Medium; Applications

Endress+Hauser Liquicap T FMI21
16 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Liquicap T FMI21
4 Endress+Hauser
Operating medium Due to the probe design, the Liquicap T FMI21 can be used as of a conductivity of 30 µs/cm. The measurement
is independent of the dc-value and the conductivity of the liquid.
L00-FMI21xxx-05-06-xx-en-000
Applications Vendor Managed Inventory
The remote interrogation of tank or silo levels via Fieldgate enables suppliers of raw materials to gather
information about the current inventories of their regular customers at any time and, for example, take this into
account in their own production planning. The Fieldgate monitor the configured level limits and automatically
trigger the next delivery as required. Here, the spectrum of possibilities ranges from a simple purchasing
requisition by e-mail through to fully automatic order processing by incorporating XML data into the planning
systems on both sides.
L00-FMI21xxx-02-00-06-en-001
10
3
10
2
10
1
1
10
10
2
10
3
10
4
30
Conductivity [µS/cm]
Liquicap T operating range
independent of the conductivity
and DK value (dielectric constant)
e.g. Hydrocarbons with
water contents below 0.1%,
petrols, oils, solvents
e.g. Hydrocarbons with
higher water contents,
demineralised water
e.g. Water-based liquids,
aqueous solutions of salts,
acids and alkalis,
aqueous dispersions and emulsions,
wastewater, electrolytes, beverages
Recommendation:
e.g. Multicap or Liquicap M
Levelflex M
Prosonic T/M
Micropilot M
Not suitable for Liquicap T
Fieldgate
FXA320
FMI21
Fieldgate
FXA320
Solar box with
FXA320
Solar panel
GSM
Vendor
Client A
Analog
4...20 mA
Analog
4...20 mA
Ethernet
Client B
Limit
switch
Telephone

Other manuals for Endress+Hauser Liquicap T FMI21

Related product manuals