Liquicap T FMI21
Endress+Hauser 7
Cable specifications Use usual commercial two-core or multi-core cable (25 per core).
Cable cross-section (incl. ferrule): max. 2.5 mm
2
!
Note!
Use a screened cable in the event of strong, electromagnetic EMC condition.
Performance characteristics with installed electronic insert
Reference operating
conditions
Ambient temperature: 23 °C
Medium temperature: 23 °C
Medium viscosity: medium must expose probe again (drain off < 2000 cst)
Atmospheric pressure
Probe installation: vertical from above
Maximum measured error 1 % of full scale value (active rod probe)
Repeatability 0.25 % of full scale value (range 0...2000 pF)
Start-up settling time < 2 s (stable measured value after switch-on process)
Influence of ambient
temperature
< 0.01 %/K (–40°C...+70 °C) probe length 1 m
Integration time τ =1 s (fixed)
The integration time affects the speed at which the display and the current output react to changes in the fluid
level.
Factory calibration
L00-FMI21xxx-15-05-xx-en-000
Factory calibration: medium conductivity 30 µS/cm
Calibration accuracy 100 % max. -5 mm; 0 % max. -5 mm
In an installed state, recalibration is only necessary if
The probe rods have been shortened
The 0 % and 100 % value should be adjusted to suit customer specifications
The electronic insert was changed
100 %
0%
Factory calibration
Recalibration
necessary
Shortening