Installation Liquiphant FTL51B
14 Endress+Hauser
5.1.5 Support the device
A0031874
 12 Support in the event of dynamic load
Support the device in the event of severe dynamic load. Maximum lateral loading capacity
of the pipe extensions and sensors: 75 Nm (55 lbf ft).
5.1.6 Weld-in adapter with leakage hole
316L
A0039230
 13 Weld-in adapter with leakage hole
Weld in the welding neck in such a way that the leakage hole is pointing downwards. This
enables any leaks to be detected quickly.
5.2 Mounting the measuring device
5.2.1 Required tools
Open-ended wrench for sensor installation
Screwdriver for electrical connection