EasyManua.ls Logo

Endress+Hauser Liquiphant FTL51B - Commissioning; Function Check

Endress+Hauser Liquiphant FTL51B
60 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Liquiphant FTL51B Commissioning
Endress+Hauser 37
Technical data
Approval: intrinsically safe Ex ia, IS or ec/ic
NAMUR electronics (electronic insert FEL68):
For energy-related reasons, the Bluetooth module VU121 requires a special battery
when operated with the 2-wire NAMUR electronics. The service life of the Bluetooth
module without replacing the battery is at least 5 years with a maximum of 60
downloads of complete datasets (at ambient temperatures between
10 to 40 °C (50 to 104 °F)).
Maximum free-field range 50 m (165 ft)
Operation radius with intervisibility 10 m (33 ft) around the device
For documentation on radio approvals, see the Endress+Hauser website:
www.endress.com → Downloads.
7.2 LED module VU120 (optional)
The LED module lights up very brightly and can be easily identified from a distance. It can
be connected to the following electronic inserts: FEL62, FEL64, FEL64 DC
Depending on either the MAX/MIN setting, the sensor status can be identified from the
LEDs; displayed in the colors green, yellow and red.
The three colors of the LEDs flash one after another as a chaser light during the functional
test.
7.2.1 Configuration and sensor status
LED shows a green light
MAX, sensor status: Not covered
MIN, sensor status: Covered
LED shows a yellow light
MAX, sensor status: Covered
MIN, sensor status: Not covered
LED shows a red light
Alarm
LED flashes red
Warning
FEL62 overload/short-circuit or no communication
The three colors of the LEDs flash one after another as a chaser light during the functional
test.
8 Commissioning
8.1 Function check
Before commissioning the measuring point, check whether the post-installation and post-
connection checks have been performed:
"Post-installation check" checklist →  16
"Post-connection check" checklist →  31