Technical data Liquiphant FTL51B
56 Endress+Hauser
14.3.9 Mechanical load
Lateral loading capacity
 Special mounting instructions
14.3.10 Electromagnetic compatibility
Electromagnetic compatibility as per EN 61326 series and NAMUR recommendation
EMC (NE21).
The requirements of EN 61326-3-1 for the safety function (SIL) are fulfilled.
Details are available in the supplementary Functional Safety Manual.
14.4 Process
14.4.1 Process temperature range
–50 to +150 °C (–58 to +302 °F)
Pay attention to the pressure and temperature dependence (see the "Sensor process
pressure range" section )
14.4.2 Thermal shock
≤ 120 K/s
14.4.3 Process pressure range
PN
T
p
+150
+302
0
32
-50
-58
100
64
1450
928
032
[bar]
[°C]
[psi]
[°F]
1
A0038268
 56 Process temperature FTL51B
1 Permitted pressure rating if the "100 bar (1 450 psi)" option is selected. For exceptions, see the "Process
connections" section. Canadian CRN approval: more details on the maximum pressure values are available in
the download area of the product page under "www.endress.com".