Liquiphant M FTL51C
Endress+Hauser 27
!
Note!
The switch points of the Liquiphant M are at other positions to those of the previous version
Liquiphant II.
Bushings: temperature spacer, pressure tight feed-through
Process connections
Any length L:
148 mm to 3000 mm (6 in to 115 in) for synthetic coating
148 mm to 1200 mm (6 in to 48 in) for enamel coating
!
Note!
The switch points of the Liquiphant M are at other positions to those of the previous version
Liquiphant II.
Special length "L II": 115 mm (4.5 in)
If vertically mounted from above, the same switch point as for the Liquiphant II
FTL360, FTL365, FDL30, FDL35
Weights See Ordering information ä 33.
Temperature spacer
Provides sealed insulation for the vessel
and normal ambient temperatures
for the housing.
L00-FTL5xxxx-06-05-xx-xx-091
Pressure tight feed-through
Protects the housing from vessel pressures
up to 40 bar (580 psi)
if the sensor is damaged.
Provides sealed insulation for the vessel
and normal ambient temperatures
for the housing.
Process connection Dimensions Accessories Pressure
Temperature
Flanges:
Synthetic coating
ANSI B16.5 (RF)
EN 1092-1 (Form A)
JIS B 2220 (RF)
Enamel coating
ANSI B16.5 (RF)
EN 1092-1 (Form B)
JIS B 2220 (RF)
A#
#
B#
#
C##
K#
#
L00-FTL51Cxx-06-05-xx-xx-024
For synthetic
coating:
seal made of PTFE
supplied
In event of enamel
coating:
Seal provided by the
customer
See nominal pressure
of flange, however
For ECTFE:
For PFA
(Edlon*):
For enamel:
40 bar
120 °C
40 bar
150 °C
25 bar
150 °C
*) FDA-compliant material in accordance with 21 CFR Part 177.1550/2600
Sensor length L
for FTL51C
L00-FTL5xxxx-06-05-xx-xx-018
+ 140
(5.51)
Isolation
≥115
(≥4.53)
mm (in)
ø24,8
L
mm (in)
(0.98)