M
Morgan EvansSep 8, 2025
Why is the measured value incorrect on my Endress+Hauser Radar?
- DDavid WardSep 8, 2025
If the measured value on your Endress+Hauser Radar is incorrect, check and adjust the Empty calibration parameter and the Full calibration parameter. If necessary, check and adjust linearization. For measurements in bypasses or stilling wells, select the correct Tank type (Bypass / pipe) and enter the correct diameter in the Tube diameter parameter. You may also need to enter the correct value in the Level correction parameter.