Nivector FTI26
12 Endress+Hauser
Short-circuit protection
• Overload protection/short-circuit protection at I > 200 mA
• IO-Link communication: 105 mA each if both switch outputs are active
Intelligent monitoring: Testing for overload at intervals of approx. 1.5 s; normal operation resumes
once the overload/short-circuit has been rectified.
Process
Process temperature range
–20 to +80 °C (–4 to +176 °F)
For Ex devices: –20 to +75 °C (–4 to +167 °F)
Process pressure range
–1 to +6 bar (–14.5 to +87 psi)
Process fluid
Powdery and fine-grained bulk solids
• Grain size < 10 mm
• Dielectric constant 1.3
• Default values: ε
r
> 1.6 with protector, ε
r
> 2.0 without protector