Proline Promag 50
irin
Endress+Hauser 37
3.2.2 Promag H connecting cable
Connecting cable termination
Terminate the signal and coil current cables as shown in the figure below (Detail A).
Fit the fine-wire cores with cable end ferrules (Detail B).
Signal cable termination
Make sure that the cable end ferrules do not touch the wire shields on the sensor side! Minimum
distance = 1 mm (0.04 in), exception "GND" = green cable.
Coil current cable termination
Insulate one core of the three-core cable at the level of the core reinforcement; you only require
two cores for the connection.
Sensor Transmitter
mm (inch)
a = 80 (3.15)
b = 15 (0.59)
c = 17 (0.67)
d = 8 (0.31)
e = ≥ 1 (≥ 0.04)
a0008173
mm (inch)
a = 80 (3.15)
b = 50 (1.97)
c = 17 (0.67)
d = 8 (0.31)
m
= Cable end ferrules, red, ∅ 1.0 mm (0.04");
n
= Cable end ferrules, white, ∅ 0.5 mm (0.02")
Sensor Transmitter
mm (inch)
a = 70 (2.76)
b = 40 (1.57)
c = 15 (0.59)
d = 8 (0.31)
a0008175
mm (inch)
a = 70 (2.76)
b = 50 (1.97)
c = 10 (0.39)
d = 8 (0.31)
m
= Cable end ferrules, red, ∅ 1.0 mm (0.04");
n
= Cable end ferrules, white, ∅ 0.5 mm (0.02")