Hardware settings Proline Promag 50
42 Endress+Hauser
PROFIBUS
A0007552
Device address range: 0 to 126
Factory setting: 126
a. Miniature switches for device address
Example shown:
1+16+32 = device address 49
b. Miniature switches for the address mode (method
of addressing):
– OFF (factory setting) = software addressing via
local operation/operating program
– ON = hardware addressing via miniature
switches
c. Miniature switches not assigned.