0NEPSI Zone 1, Zone 21 Proline Promag 53
6 Endress+Hauser
Type (compact/remote; Item No. 13 in type code)
Type of cable entry (Item No. 15 in type code)
Display, operation, power supply (Item No. 16 in type code)
Outputs/inputs (Item No. 18 in type code)
!
Note!
A detailed explanation of these values, regarding the available outputs and inputs, as well as a description of
the associated terminal assignments and connection data can be found from Page 11 onwards.
Temperature table
compact version
Max. medium temperature [°C] for T1 ~ T6 in relation to the maximum ambient temperature T
a
1)
with polyurethane lining;
2)
with hard rubber lining
1)
with PFA lining;
2)
with PTFE lining
The minimum ambient temperature is –20 °C.
A version for ambient temperatures up to –40 °C is optionally available
* Type Min. ambient temperature T
a min
A, P, U
Compact
–20 °C
V, 6 –40 °C
G, N, T, W
Remote
–20 °C
7, 8 –40 °C
* Thread form
A M20 × 1.5
BNPT ½"
CG ½"
* Thread form
A, C, E, G, P, R, 0, 2, 4, 7 85 – 260 V AC
B, D, F, H, Q, S, 1, 3, 5, 8 20 – 55 V AC/16 – 62 V DC
* Type of protection
A, B, C, D, J, K, L, M, N, P, Q, V, 2, 4, 5, 7 non Ex ia (intrinsically safe) input /output
F, G, S, T Ex ia (intrinsically safe) input /output
DN T
a
T6
(85 °C)
T5
(100 °C)
T4
(135 °C)
T3
(200 °C)
T2
(300 °C)
T1
(450 °C)
Promag 5*W**-************
25 to 1200
1)
+50 °C 50 50 50 50 50 50
50 to 2000
2)
+50 °C 80 80 80 80 80 80
DN T
a
T6
(85 °C)
T5
(100 °C)
T4
(135 °C)
T3
(200 °C)
T2
(300 °C)
T1
(450 °C)
Promag 5*P**-************
25 to 200
1)
+40 °C 80 95 130
150 150 150
15 to 600
2)
130 130 130
25 to 200
1)
+45 °C 80 95 130 130 130 130
15 to 600
2)
25 to 200
1)
+50 °C 80 95 95 95 95 95
15 to 600
2)
DN T
a
T6
(85 °C)
T5
(100 °C)
T4
(135 °C)
T3
(200 °C)
T2
(300 °C)
T1
(450 °C)
Promag 5*H**-************ 2 to 150
+40 °C 80 95 130 150 150 150
+45 °C 80 95 130 130 130 130
+50 °C 80 95 95 95 95 95