Electrical connection Proline Promag L 400
32 Endress+Hauser
!
WARNING
Housing degree of protection may be voided due to insufficient sealing of the housing.
Screw in the screw without using any lubricant. The threads on the cover are coated with a
dry lubricant.
Tightening torques for plastic housing
Housing cover fixing screw 1.3 Nm
Cable entry 4.5 to 5 Nm
Ground terminal 2.5 Nm
7.2.1 Connecting the transmitter
Supply voltage connection, 0-20 mA/4-20 mA HART and additional outputs/inputs
TX 20
4 x
1.
2.
3.
26
27
24
25
23 20
21
22
+ - + -
+ - + -
1
2
L+/L
L-/N
6.
4.
5.
10 (0.4)
A0017268
Connect the cable in accordance with the terminal assignment (→  27). For supply
voltage: open the shock protection cover.
For HART communication: When connecting the cable shielding to the ground terminal,
observe the grounding concept of the facility.