Electrical connection Proline Promag L 400
34 Endress+Hauser
Connecting the supply voltage and EtherNet/IP
PH 2
4 x
1.
2.
3.
4.
5.
10 (0.4)
1
2
L+/L
L-/N
6.
A0021356
Connect the cable in accordance with the terminal assignment (→  27). For supply
voltage: open the shock protection cover.
7.2.2 Connecting the remote version
!
WARNING
Risk of damaging the electronic components!
Ground the remote version and in doing so connect the sensor and transmitter to the same
potential equalization.
Only connect the sensor to a transmitter with the same serial number.
Ground the connection housing of the sensor via the external screw terminal.
The following procedure (in the action sequence given) is recommended for the remote
version:
1. Mount the transmitter and sensor.
2. Connect the connecting cable.
3. Connect the transmitter.