Proline Promag W 400 Electrical connection
Endress+Hauser 27
7.1.6 Requirements for the supply unit
Supply voltage
Transmitter
Order code for "Power supply" Terminal voltage Frequency range
Option L
AC100 to 240 V 50/ 60 Hz, ±4 Hz
AC/DC24 V 50/ 60 Hz, ±4 Hz
7.1.7 Preparing the measuring device
1. Remove dummy plug if present.
2. If measuring device is delivered with cable glands:
Observe cable specification →  21.
7.1.8 Preparing the connecting cable for the remote version
When terminating the connecting cable, pay attention to the following points:
In the case of electrode cables, make sure that the ferrules do not touch the core shields on
the sensor side. Minimum distance = 1 mm (exception: green “GND” cable)
In the case of coil current cables, insulate one core of the three-core wire at the level of the
core reinforcement. You only require two cores for the connection.
Fit the fine-wire cores with ferrules.