Proline Promag W 400 Operation options
Endress+Hauser 47
Measuring device
Web server Web server must be enabled; factory setting: ON
IP address If the IP address of the device is not known:
• The IP address can be read out via local operation:
"Diagnostics" menu → Device information → IP address
• Communication with the Web server can be established via the standard IP
address 192.168.1.212.
The DHCP function is enabled in the device at the factory, i.e. the device
expects an IP address to be assigned by the network. This function can be
disabled and the device can be set to the standard IP address 192.168.1.212:
set top DIP switch No. 2 from OFF → ON.
A0023353
• Once the DIP switch has been activated, the device must be restarted
before the device uses the standard IP address.
• If the standard IP address (top DIP switch No. 2 = ON) is used, there is
no connection to the EtherNet/IP network.
8.3.3 Establishing a connection
Configuring the Internet protocol of the computer
IP address 192.168.1.XXX; for XXX all numerical values except: 0, 212 and 255 → e.g.
192.168.1.213
Subnet mask 255.255.255.0
Default gateway 192.168.1.212 or leave cells empty
1. Switch on the measuring device and connect to the computer via the cable .
2. If a 2nd network card is not used: all the applications on the notebook should be closed,
or all the applications that require the Internet or network, such as e-mail, SAP
applications, Internet or Windows Explorer, i.e. close all open Internet browsers.
3. Configure the properties of the Internet protocol (TCP/IP) as defined in the table above.