Commissioning Proline Prowirl 73
28 Endress+Hauser
Option output type
Only the output (current output or pulse/status output) not yet configured in the current
Quick Setup is offered for selection after the first cycle.
The "YES" option appears as long as a free output is still available. "NO" is the only option
displayed when no further outputs are available.
Automatic configuration of display
Selection Output type
Current output Frequency output Quit
Assign current Selection frequency Selection pulse Selection status QS Commisioning
Current range
Value 4 mA Assign frequency Assign pulse Assign status
Value 20 mA End value frequency Pulse value Switch on point
Time constant Value f low Pulse width Switch off point
Failsafe mode Value f high Output signal Time constant
Output signal Failsafe mode
Time constant
Failsafe mode
Configuration another output?
YES Option output type / NO Automatically configuration display
Automatic configuration of display?
YES NO
Automatic parameterization
of the display
QS Commisioning or
called up the group flow computer (only by selection: Real gas, Natural gas NX-19)