Installation Proline Prowirl R 200
12 Endress+Hauser
6 Installation
6.1
Installation conditions
No special measures such as supports are necessary. External forces are absorbed by
the construction of the device.
6.1.1 Mounting position
Mounting location
A0015543
Orientation
Vortex meters require a fully developed flow profile as a prerequisite for correct volume
flow measurement.
The direction of the arrow on the sensor nameplate helps you to install the sensor
according to the flow direction.
The device can basically be installed in any orientation in the pipe. However, note the
following points:
Orientation Compact version Remote version
A Vertical orientation
A0015545

1)

B Horizontal orientation,
transmitter head up
A0015589

2) 3)
