Commissioning Proline Prowirl R 200
52 Endress+Hauser
Defining the access code via local display
1. Navigate to the "Enter access code" parameter.
2. Define a max. 4-digit numeric code as an access code.
3. Enter the access code again to confirm the code.
The
-symbol appears in front of all write-protected parameters.
The device automatically locks the write-protected parameters again if a key is not
pressed for 10 minutes in the navigation and editing view. The device locks the write-
protected parameters automatically after
60 s if the user skips back to the operational
display mode from the navigation and editing view.
If write access is activated via access code, it can be also be deactivated only via
the access code (→ 
47).
The user role with which the user is currently logged on via the local display
(→  46) is indicated by the "Access status display" parameter. Navigation
path: Operation → Access status display
10.6.2 Write protection via write protection switch
Unlike write protection via user-specific access code, this allows write access to the
entire operating menu - other than the "Contrast display" parameter - to be locked.
The parameter values are now read only and cannot be edited any more (exception
"Contrast display" parameter):
Via local display
Via service interface (CDI)
Via HART protocol
Via PROFIBUS PA protocol