3 Safety instructions
Only use original parts from Endress+Hauser.
Comply with national regulations governing mounting, electrical installation,
commissioning, maintenance and device conversion procedures.
The specialized technical staff involved in the mounting, electrical installation,
commissioning, maintenance and modification of the measuring devices must meet the
following requirements:
Specialized technical staff must be trained in instrument safety.
They must be familiar with the individual operating conditions of the devices.
In the case of Ex-certified measuring devices, they must also be trained in explosion
protection.
The measuring device is energized! Risk of fatal injury from electric shock. Open the
measuring device only when the device is de-energized.
When using Ex-certified measuring devices: open device only when in a de-energized state
(allow 10 minutes to elapse after switching off the power supply) or in environments which
do not have a potentially explosive atmosphere.
For measuring devices in safety-related applications in accordance with IEC 61508 or
IEC 61511: following modification, re-commission the device in accordance with the
Operating Instructions. Document the modification.
Before removing the device: set the process to a safe state and purge the pipe of dangerous
process substances.
Hot surfaces! Risk of injury! Before commencing work: allow the system and measuring
device to cool down to a touchable temperature.
In the case of devices in custody transfer, the custody transfer status no longer applies once
the seal has been removed.
The Operating Instructions for the device must be followed.
Risk of damaging the electronic components! Ensure you have a working environment
protected from electrostatic discharge.
After removing the electronics compartment cover: risk of electrical shock due to missing
touch protection!
Turn the measuring device off before removing internal covers.
Only open the housing for a brief period. Avoid the penetration of foreign bodies, moisture
or contaminants.
Replace defective seals only with original seals from Endress+Hauser.
If threads are damaged or defective, the measuring device must be repaired.
Threads (e.g. of the electronics compartment cover and connection compartment cover)
must be lubricated if an abrasion-proof dry lubricant is not available. Use acid-free, non-
hardening lubricant.