Prosonic S FMU90 Flow Measurement PROFIBUS Commissioning
Endress+Hauser 37
5.5 Basic setup
After the first setup the main screen appears (flow or level).
However, the displayed values do not yet correspond to the real flows or level before you have
performed the basic setup. To do so, enter the main menu by pressing "Menu" (right key). Select
the "flow" or the "level" submenu.
The "flow" submenu is used for the calibration for the following tasks:
Flow measurements (1 or 2 channels)
Back water alarm
•Flow counters
The structure of the submenu depends on the selected operating mode
4)
:
Always start by calibrating the first flow channel ("flow 1" submenu). Thereafter, you can
calibrate the following as required:
Second flow channel ("flow 2" submenu)
Backwater detection ("backwater" submenu)
•Flow counters ("flow counter" submenu)
4) The operating mode is selected during the first setup. Nevertheless, it can be changed at any time if required ("device
properties" menu, "operating params" submenu, "operating mode" parameter set).
Main menu Submenu
Operating modes"level+flow", "flow"
Main menu flow Flow Flow 1
Flow 2
Basic setup
Flow counter
Operating mode "flow+backwater"
Main menu flow Flow 1+backwater Flow
Backwater
Basic setup
Flow counter