Commissioning Prosonic S FMU90 Flow Measurement PROFIBUS
38 Endress+Hauser
The "level" submenu is used for the calibration of level measurement (1 or 2 channels)
5.6 Overview basic setup for a flow measurement
The following table gives an overview of the basic setup for flow measurement. Detailed
information on the parameters can be found in the Operating Instructions BA00293F.
Step Parameter Set Parameter Remarks
1 Main menu level Level (LVL)1
Level (LVL)2
1)
1) The selection "Level (LVL) 2" is only available for instruments with 2 sensor inputs or 2 current outputs.
Select the "level" submenu. In the following submenu
select the level channel you are going to calibrate.
2 Main menu level
level (LVL)1
Basic setup In the following submenu select "basic setup".
This submenu contains all parameters needed for the
basic setup.
Step Parameter Set Parameter Remarks
1 Open the "flow 1" or "flow 2" submenu
2 Open the "basic setup" submenu.
3 Sensor selection
(N = 1 or 2)
Input Allocate a sensor to the channel.
Sensor selection Specify the type of sensor ("automatic" for FDU9x).
Detected Only available for "sensor selection" = "automatic"; indicates
the detected type of sensor
4 Flow N linearization
(N = 1 or 2)
Type Select type of linearizationn
1)
:
"Flume/weir" (for the pre-programmed flumes and weirs)
"Table" (to enter a lineariztion table manually)
"Formula" (for the flow formula Q = C (h
α
+ γh
β
))
(Details ä 41 "type")
Flow unit Select the flow unit (Details ä 41 "flow unit")
Curve Select the type of flume or weir; (only present for "type" =
"flume/weir"); A second page appears in which the size of
the respective flume and weir must be selected
(Details Operating Instructions BA00293F)
Edit Used to enter, change or delete a linearization table, (only
available for "type" = "table"), (Details ä 42 "edit")
Status table Enables or disables the linearization table,
(only available for "type" = table")
Alpha Specify the value of the parameter α,
(only available for "type" = "formula"