Prosonic S FMU90 Flow Measurement PROFIBUS Commissioning
Endress+Hauser 47
5.7.1 Application parameters
Parameter Selection/Remarks
"tank shape" Use this parameter to specify the tank shape of your application:
Selection:
• dome ceiling (→ A)
• horizontal cyl. (→ B)
• bypass (→ C)
• stilling well (ultrasonic guide pipe) (→ C)
• no ceiling (→ D)
• sphere (→ E)
• flat ceiling (→ F)
L00-FMU30KAx-14-00-06-xx-001
"medium property" Use this parameter to specify the type of medium.
You have the following options:
• Unknown (e.g. pasty media such as greases, creams, gels etc.)
•Liquid
• Solid, grain size < 4 mm (fine)
• Solid, grain size > 4 mm (coarse)
If the medium does not fit into one of the groups, select "Unknown".
"process conditions" Use this parameter to specify the process conditions of your application. The filters of the
signal evaluation are automatically adjusted to the selected conditions.
For this function, you have the following options → see the table: