EasyManua.ls Logo

Endress+Hauser Prosonic S FMU90 - Terminal Assignment

Endress+Hauser Prosonic S FMU90
36 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Prosonic S FMU90
Endress + Hauser 13
Terminal assignment Pluggable spring-force terminals for connection of the cables are supplied in the terminal compartment. Rigid
conductors or flexible conductors with cable sleeve can directly be inserted and are contacted automatically.
The terminal configuration depends on the instrument version ordered. There is a basic terminal area, which
is present in every instrument version. Additonal optional terminal areas are only present if the respective
option has been selected in the product structure.
L00-FMU90xxx-04-00-00-xx-001
Terminals of the Prosonic S; the terminals depicted in grey are not present in every instrument version.
A: Basic terminal area; B-E: Optional terminal areas (present if the respective option has been selected in the product
structure)
!
Note!
The depicted switching states of the relays refer to the de-energized state.
Conductor cross section 0,2 mm
2
- 2,5 mm
2
Cable and sleeve cross section 0,25 mm
2
- 2,5 mm
2
min. stripping length 10 mm
Terminal area present for the following instrument versions
Basic area A for all versions
Optional areas
B
for instrument versions with 2 sensor inputs and/or 2 analogue outputs
(FMU90 - *****2****** and/or FMU90 - *******2****)
C
for instrument versions with 3 or 6 relays
(FMU90 - ******3***** oder FMU90 - ******6*****)
D
for instruments with external switch inputs and external temperature input
(FMU90 - ********B***)
E
for instrument versions with PROFIBUS DP interface
(FMU90 - *******3****)
FDU-
Sensor
0/4...20mA
1
42
41
Relay
5552
5451 53
50
3
Address
Term.
DP
off
on
off
on
SW
HW
1
2
3
4
5
6
7
8
1
2
3
4
A(N)
66
B(P)
65
Display
POWER
HART
0/4…20mA
Sync
Fuse
I
1
FDU-
Sensor
RD
11
BK
10
YE
9
40
39
5
4
67
8
1
1
Service
Relay
321
A
B C E
2
58
5756
4
61
61
60
59
59
5
64
63
62
62
6
RD
11
BK
10
YE
9
2
RD
14
BK
13
YE
12
I
2
FDU-
Sensor
DigIn
76
73
7572 74
71
2
D
1
79
79
7877
3
82
81
80
4
85
8483
Temp.

Table of Contents

Other manuals for Endress+Hauser Prosonic S FMU90