Prosonic T Mounting
Endress+Hauser 11
2.5 Installation hint
2.6 Installation check
After installing the device, carry out the following checks:
• Is the device damaged (visual inspection)?
• Does the device correspond to the measuring point specifications for process temperature,
process pressure, ambient temperature, measuring range etc.
• If available: Are the measuring point number and labelling correct (visual inspection)?
• Is the measuring device sufficiently protected against precipitation and direct sunlight?
• Are the cable glands tightened correctly?
• After aligning the housing, check the process seal at the nozzle or flange.
Screw the instrument at the screw-in
piece using an 65AF spanner.
Maximum torque: 7 Nm (5.16 lbf ft).
L00-FMU30xxx-17-00-00-en-009
65
AF65
max torque
7 Nm (5.16 lbf ft)
Caution!
Use only the screw-in
piece to screw in
the instrument