Loading...
Prosonic T Mounting
Endress+Hauser 11
2.5 Installation hint
2.6 Installation check
After installing the device, carry out the following checks:
Is the device damaged (visual inspection)?
Does the device correspond to the measuring point specifications for process temperature,
process pressure, ambient temperature, measuring range etc.
If available: Are the measuring point number and labelling correct (visual inspection)?
Is the measuring device sufficiently protected against precipitation and direct sunlight?
Are the cable glands tightened correctly?
After aligning the housing, check the process seal at the nozzle or flange.
Screw the instrument at the screw-in
piece using an 65AF spanner.
Maximum torque: 7 Nm (5.16 lbf ft).
L00-FMU30xxx-17-00-00-en-009
65
AF65
max torque
7 Nm (5.16 lbf ft)
Caution!
Use only the screw-in
piece to screw in
the instrument